With nearly sixty years experience in the design of spectrophotometers, JASCO offers a complete range of UV-Visible/NIR spectrophotometers. The V-700 series comprises of five distinct models covering the entire spectrum from the Far UV (187nm) right through to the NIR (3200nm). Each model has been designed to meet the exacting requirements of any analytical laboratory.
| Specification | V-730 | V-750 | V-760 | V-770 | V-780 |
|---|---|---|---|---|---|
| Optical system | Rowland off-circle arrangement Single monochromator Double beam type |
Czerny-Turner mount Single monochromator Fully symmetrical double beam type |
Czerny-Turner mount Double monochromator Fully symmetrical double beam type |
Czerny-Turner mount Single monochromator Fully symmetrical double beam type |
Czerny-Turner mount Single monochromator Fully symmetrical double beam type |
| Light source | Halogen lamp, deuterium lamp | Halogen lamp, deuterium lamp | Halogen lamp, deuterium lamp | Halogen lamp, deuterium lamp | Halogen lamp, deuterium lamp |
| Wavelength range | 190 to 1,100 nm | 190 to 900 nm | 187 to 900 nm | 190 to 2,700 nm (3,200 nm option) | 190 to 1,600 nm |
| Wavelength accuracy | ±0.2 nm (at 656.1 nm) | ±0.2 nm (at 656.1 nm) | ±0.1 nm (at 656.1 nm) | ±0.3 nm (at 656.1 nm) ±1.5 nm (at 1,312.2 nm) |
±0.3 nm (at 656.1 nm) ±1.0 nm (at 1,312.2 nm) |
| Wavelength repeatability | ±0.1 nm | ±0.05 nm | ±0.05 nm | ±0.05 nm (UV-Vis) ±0.2 nm (NIR) |
±0.05 nm (UV-Vis) ±0.1 nm (NIR) |
| Spectral bandwidth (SBW) | 1 nm | 0.1, 0.2, 0.5, 1, 2, 5, 10 nm L2, L5, L10 nm (low stray light mode) M1, M2 nm (micro cell mode) |
0.1, 0.2, 0.5, 1, 2, 5, 10 nm L2, L5, L10 nm (low stray light mode) M1, M2 nm (micro cell mode) |
UV-Vis: 0.1, 0.2, 0.5, 1, 2, 5, 10 nm L2, L5, L10 nm (low stray light mode) M1, M2 nm (micro cell mode) NIR: 0.4, 0.8, 1, 2, 4, 8, 20, 40 nm L8, L20, L40 nm (low stray light mode) M4, M8 nm (micro cell mode) |
UV-Vis: 0.1, 0.2, 0.5, 1, 2, 5, 10 nm L2, L5, L10 nm (low stray light mode) M1, M2 nm (micro cell mode) NIR: 0.2, 0.4, 0.5, 1, 2, 4, 10, 20 nm L4, L10, L20 nm (low stray light mode) M2, M4 nm (micro cell mode) |
| Stray light | 1 % (198 nm, KCl 12 g/L aqueous solution) 0.02 % (220 nm, NaI 10 g/L aqueous solution) 0.02 % (340 nm, NaNO2 50 g/L aqueous solution) 0.02 % (370 nm, NaNO2 50 g/L aqueous solution) SBW: 1 nm |
1 % (198 nm, KCl 12 g/L aqueous solution) 0.005 % (220 nm, NaI 10 g/L aqueous solution) 0.005 % (340 nm, NaNO2 50 g/L aqueous solution) 0.005 % (370 nm, NaNO2 50 g/L aqueous solution) SBW: L2 nm |
1 % (198 nm, KCl 12 g/L aqueous solution) 0.00008 % (220 nm, NaI 10 g/L aqueous solution) 0.00008 % (340 nm, NaNO2 50 g/L aqueous solution) 0.00008 % (370 nm, NaNO2 50 g/L aqueous solution) SBW: L2 nm |
1 % (198 nm, KCl) 0.005 % (220 nm, NaI) 0.005 % (340 nm, NaNO2) 0.005 % (370 nm, NaNO2) SBW: L2 nm 0.04 % (1,420 nm, H2O) 0.1 % (1,690 nm, CH2Br2) SBW: L8 nm |
1 % (198 nm, KCl 12 g/L aqueous solution) 0.005 % (220 nm, NaI 10 g/L aqueous solution) 0.005 % (340 nm, NaNO2 50 g/L aqueous solution) 0.005 % (370 nm, NaNO2 50 g/L aqueous solution) SBW: L2 nm 0.04 % (1,420 nm, H2O) SBW: L4 nm |
| Photometric range | -3 to 3 Abs | -4 to 4 Abs | -4 to 6 Abs | UV-Vis: -4 to 4 Abs NIR: -3 to 3 Abs |
UV-Vis: -4 to 4 Abs NIR: -3 to 3 Abs |
| Photometric accuracy | ±0.0015 Abs (0 to 0.5 Abs) ±0.0025 Abs (0.5 to 1 Abs) ±0.3 %T Tested with NIST SRM 930D |
±0.0015 Abs (0 to 0.5 Abs) ±0.0025 Abs (0.5 to 1 Abs) ±0.3 %T Tested with NIST SRM 930D |
±0.0015 Abs (0 to 0.5 Abs) ±0.0025 Abs (0.5 to 1 Abs) ±0.3 %T Tested with NIST SRM 930D |
±0.0015 Abs (0 to 0.5 Abs) ±0.0025 Abs (0.5 to 1 Abs) ±0.3 %T Tested with NIST SRM 930D |
±0.0015 Abs (0 to 0.5 Abs) ±0.0025 Abs (0.5 to 1 Abs) ±0.3 %T Tested with NIST SRM 930D |
| Photometric repeatability | ±0.0005 Abs (0 to 0.5 Abs) ±0.0005 Abs (0.5 to 1 Abs) Tested with NIST SRM 930D |
±0.0005 Abs (0 to 0.5 Abs) ±0.0005 Abs (0.5 to 1 Abs) Tested with NIST SRM 930D |
±0.0005 Abs (0 to 0.5 Abs) ±0.0005 Abs (0.5 to 1 Abs) Tested with NIST SRM 930D |
±0.0005 Abs (0 to 0.5 Abs) ±0.0005 Abs (0.5 to 1 Abs) Tested with NIST SRM 930D |
±0.0005 Abs (0 to 0.5 Abs) ±0.0005 Abs (0.5 to 1 Abs) Tested with NIST SRM 930D |
| Scanning speed | 10 to 8,000 nm/min | 10 to 4,000 nm/min (8,000 nm/min in preview mode) | 10 to 4,000 nm/min (8,000 nm/min in preview mode) | 10 to 4,000 nm/min (8,000 nm/min in preview mode) | 10 to 4,000 nm/min (8,000 nm/min in preview mode) |
| Slew speed | 24,000 nm/min | 12,000 nm/min | 12,000 nm/min | UV-Vis: 12,000 nm/min NIR: 48,000 nm/min |
UV-Vis: 12,000 nm/min NIR: 24,000 nm/min |
| RMS noise | 0.00004 Abs (0 Abs, wavelength: 500 nm, measurement time: 60 sec, SBW: 1 nm) |
0.00003 Abs (0 Abs, wavelength: 500 nm, measurement time: 60 sec, SBW: 2 nm) |
0.00003 Abs (0 Abs, wavelength: 500 nm, measurement time: 60 sec, SBW: 2 nm) |
0.00003 Abs (0 Abs, wavelength: 500 nm, measurement time: 60 sec, SBW: 2 nm) |
0.00003 Abs (0 Abs, wavelength: 500 nm, measurement time: 60 sec, SBW: 2 nm) |
| Baseline stability | 0.0004 Abs/hour (more than one hour after turning on the source, with room temperature stabilised, wavelength: 250 nm, response: slow) |
0.0003 Abs/hour (wavelength: 250 nm, response: slow, SBW: 2 nm) |
0.0003 Abs/hour (wavelength: 250 nm, response: slow, SBW: 2 nm) |
0.0003 Abs/hour (wavelength: 250 nm, response: slow, SBW: 2 nm) |
0.0004 Abs/hour (wavelength: 250 nm, response: slow, SBW: 2 nm) |
| Baseline flatness | ±0.0005 Abs (200 to 1,000 nm) | ±0.0002 Abs (200 to 850 nm) | ±0.0003 Abs (200 to 800 nm) | ±0.0002 Abs (200 to 2,500 nm) | ±0.0002 Abs (200 to 1,600 nm) |
| Detector | Silicon photodiode | Photomultiplier tube | Photomultiplier tube | Photomultiplier tube, Peltier cooled PbS | Photomultiplier tube, Peltier cooled InGaAs photodiode |
| Standard functions | IQ accessory, start button, analogue output | IQ accessory, start button, analogue output | IQ accessory, start button, analogue output | IQ accessory, start button, analogue output | IQ accessory, start button, analogue output |
| Standard programs | Abs/%T meter, quantitative analysis, spectrum measurement, time course measurement, fixed wavelength measurement, validation, daily maintenance, two wavelength time course measurement | Abs/%T meter, quantitative analysis, spectrum measurement, time course measurement, fixed wavelength measurement, validation, daily check, two wavelength time course measurement | Abs/%T meter, quantitative analysis, spectrum measurement, time course measurement, fixed wavelength measurement, validation, daily check, two wavelength time course measurement | Abs/%T meter, quantitative analysis, spectrum measurement, time course measurement, fixed wavelength measurement, validation, daily check, two wavelength time course measurement | Abs/%T meter, quantitative analysis, spectrum measurement, time course measurement, fixed wavelength measurement, validation, daily check, two wavelength time course measurement |
| Dimensions and weight | 486 (W) x 441 (D) x 216 (H) mm, 15 kg | 460 (W) x 602 (D) x 268 (H) mm, 27 kg | 460 (W) x 602 (D) x 268 (H) mm, 29 kg | 460 (W) x 602 (D) x 268 (H) mm, 29 kg | 460 (W) x 602 (D) x 268 (H) mm, 29 kg |
| Power requirements | 120 VA | 150 VA | 150 VA | 150 VA | 150 VA |
| Installation requirements | Room temperature: 15 to 30 °C, humidity: below 85 % | Room temperature: 15 to 30 °C, humidity: below 85 % | Room temperature: 15 to 30 °C, humidity: below 85 % | Room temperature: 15 to 30 °C, humidity: below 85 % | Room temperature: 15 to 30 °C, humidity: below 85 % |